The ST5010, CV5030, and HM355S should first be valued as separable instruments. The presence of both Leica machines does not prove the TS5015 hardware required for an integrated transfer workflow.
No power-on, staining, coverslipping, motor-feed, or sectioning test is supplied. The bid model remains well below tested or dealer-supported market context and must absorb substantial packing and removal uncertainty.
- Inventory Leica racks, vessels, plumbing, coverslip/dispenser components, HM355S holder hardware, and actual loading terms before raising the ceiling.
- Require initialization and representative workflow checks for both Leica instruments plus HM355S feed and repeatable-sectioning evidence before using tested-working pricing.