Analysis snapshotSnapshot decision:REVIEW

Binding Site Optilite Type 864 Special Protein Analyzer — GSA Auction Analysis

A photo-confirmed 2019 Binding Site Optilite Type 864 is seller-reported operational, with visible robotics, fluid containers, and an accessory tote, but controller/software, complete peripherals, startup, QC, and assay performance are not independently established. The $6,000 conservative model supports REVIEW and a $1,000 blind ceiling.

Analysis snapshot · Historical — observed September 16, 2026. The scheduled close date of Sep 17, 2026, 11:59 PM UTC has passed. The price, evidence, economics, and recommendation remain exactly as recorded and are not current.

Decision at a glance

RecommendationREVIEW
Maximum hammer bid
$1,000

The auction bid ceiling supported by this analysis, if one is defensible; see costs below for buyer-premium treatment.

Why this decision

REVIEW: the exact 2019 Type 864 identity and seller-reported operation are useful, but independent functional proof, controller/software completeness, QC/service history, and realized-market evidence remain unresolved; use the $1,000 ceiling only as a blind/as-is threshold.

What to verify before bidding

  • A display-like peripheral does not establish a compatible controller, usable Optilite software, entitlement, or complete workflow.
  • Seller-reported operation does not establish robotics, fluidics, optics, QC, calibration, or assay performance.
Review all risks and unknowns

See costs and maximum bid for the displayed assumptions and buyer-premium treatment.

Snapshot auction result
Not captured
Bidders
Not captured
Blind-model resale
$6,000
Snapshot provenance and timing · observed Sep 16, 2026, 8:02 AM UTC · scheduled close Sep 17, 2026, 11:59 PM UTC

GSA Auctions sale 2-1-QSC-I-26-388, lot 004 · Brockton, Massachusetts
Snapshot observed Sep 16, 2026, 8:02 AM UTC
Scheduled close Sep 17, 2026, 11:59 PM UTC
Analysis prepared Sep 16, 2026, 8:02 AM UTC
Published September 16, 2026 · page updated September 16, 2026

Prepared and published by Oleksii Udod for LotCaliper.

View official GSA lot

Continue after this auction

This auction snapshot is historical. Continue with related research, or review a relevant published analysis whose recorded scheduled close has not passed.

These are individually prepared, time-stamped analyses with recorded scheduled closes. Verify the official auction before acting.

Auction snapshot

This page preserves the September 16, 2026 GSA open-auction observation for sale 2-1-QSC-I-26-388, lot 004, item 377319. No numeric high bid or bidder count was displayed; blank fields are not a $0 result.

Seller-reported operation and a complete-looking analyzer do not establish a transferable controller/software environment or independently demonstrated clinical-laboratory performance.

Review and correction history

Independent review completed. No corrections recorded.

View review and correction details

Independent review completed

Status: Approved · Reviewed September 16, 2026

A source-first second-pass review preserved seller-reported operation without promoting it to independently verified function.

How independent review works
How evidence is classified

Seller facts, photo-derived observations, manufacturer context, unknowns, and LotCaliper assumptions remain explicitly separated.

Source Confirmed

Seller-confirmed

  • Optilite Type 864 identity and 2019 manufacture are photo-confirmed; seller reports the analyzer operational.

Photo Derived

Photo-derived

  • Internal robotics, front fluid-container positions, a reservoir, cabling, and a display-like tote peripheral are visible.

Not Established

Not established

  • Compatible controller/software, complete peripherals, initialization, robotics, fluidics, optics, QC, calibration, assay performance, service/decontamination history, and freight cost.

Manufacturer Context

Independent context

  • Thermo Fisher continues to publish Optilite special-protein platform context, which does not establish support or function for this unit.

Lotcaliper Assumption

LotCaliper assumptions

  • $6,000 conservative resale after inventory and basic diagnostics, without assigning tested clinical performance, validated QC, or transferable software value.
  • $1,000 blind maximum hammer after selling friction, logistics, diagnostics, configuration, uncertainty, and required-profit reserves.
Structured photo observations (3)

These are bounded observations from the preserved photographs. They do not establish functional operation or seller inclusion beyond the corresponding source evidence.

Model · Confirmed

Clinical Chemistry Analyzer Type 864, manufactured 2019

Photo confidence: High

Module Accessory · Confirmed

Internal robotic/sample-handling assemblies visibly present

Photo confidence: Medium High

Module Accessory · Probable

Display-like peripheral in accessory tote

Photo confidence: Medium

Equipment identity

Laboratory · quantity 1

Binding Site Optilite Type 864 special protein analyzer, manufactured 2019

Identity confidence: High

Identity evidence
  • The auction record and visible manufacturer label identify an Optilite Type 864 manufactured in 2019.

    GSA Auctions · Image · observed Sep 16, 2026, 8:02 AM UTC

Configuration

Internal robotic/sample-handling assemblies and populated front fluid-container positions

Confirmed · Medium High confidence

Configuration evidence
  • Robotic/sample-handling assemblies and multiple front fluid-container positions are visible through the open access area.

    GSA Auctions · Image · observed Sep 16, 2026, 8:02 AM UTC

Accessory tote with reservoir, cabling, and display-like peripheral

Possible · Medium confidence

Configuration evidence
  • A reservoir, cables, and a display-like peripheral are visible in the tote, but exact identity and completeness are not readable.

    GSA Auctions · Image · observed Sep 16, 2026, 8:02 AM UTC

Compatible controller PC, usable Optilite software environment, licenses, LIS configuration, and complete peripherals

Not established · no source evidence establishes presence or absence

Configuration evidence
  • Visible peripherals do not establish a compatible controller, usable software, entitlement, interfaces, or complete system workflow.

    GSA Auctions · Image · observed Sep 16, 2026, 8:02 AM UTC

Condition

Seller Reported Operational

The seller reports the analyzer operational; that statement is not independent startup, QC, or assay-performance evidence.

Evidence confidence: Medium

Condition evidence
  • Seller condition text reports operational status while retaining missing-parts and repair caveats.

    GSA Auctions · Text · observed Sep 16, 2026, 8:02 AM UTC

Untested

No independently demonstrated initialization, robotics, fluidics, temperature control, optics, barcode handling, QC, calibration, or application-level result is preserved.

Evidence confidence: High

Condition evidence
  • Substantial exterior and internal assemblies are visible, but no operating screen, test sequence, or result is supplied.

    GSA Auctions · Image · observed Sep 16, 2026, 8:02 AM UTC

Market evidence

Market evidence strength

Medium

Public auditability

Full

Evidence recency

Mixed

Current and archived exact-model asking evidence establishes a specialist price surface, but no qualified exact-model realized sale supports a tested-working resale claim.

Model matchEvidence typeTreatmentCount
Exact ModelAskingUsed In Valuation3
Exact ModelEnded Non RealizedExcluded1
SourceEvidenceObservedPriceRelevance and treatment
NorthWest Supply / DOTmed Optilite

Dealer

Asking

Exact Model

September 16, 2026$9,000

2017 used Type 864 rated good/working with dealer return and limited parts-warranty terms.

Used In Valuation

Bimedis archived Optilite

Marketplace

Asking

Exact Model

September 16, 2026$6,999

Archived exact-model used ask with thin configuration and condition detail.

Used In Valuation

Bimedis archived 2017 Optilite

Marketplace

Asking

Exact Model

September 16, 2026$11,000

Archived exact-model 2017 used/good ask; stronger stated condition and no proof of sale.

Used In Valuation

Sensitivity and robustness

Requires Verification

The scenarios show potentially workable economics, but the stated verification gate must be resolved before bidding responsibly.

Verified complete system

Max hammer
$3,500
Gross resale model
$7,000–$8,000

A conditional ceiling near $3,500 requires a complete controller/software environment, demonstrated basic function, lower diagnostic/configuration reserves, and a stronger resale basis.

Costs and maximum bid

Transaction / operating costAmountBasis
Selling Fees$900LotCaliper assumption: 15% of modeled gross resale.
Shipping$700LotCaliper pickup, packing, rigging, and inbound-logistics reserve; not a carrier or GSA quote.
Testing Repair Reserve$600LotCaliper functional-test and diagnostic reserve.
Risk Buffer$800Unresolved controller/software, robotics, fluidics, optics, QC, repair, and specialist-liquidity risk reserve.
Other$750LotCaliper missing-component, configuration, or recovery reserve; not a seller or service quote.
Buyer premium
Not modeled
Return requirements
$500 minimum profit · 30% minimum ROI
Computed bid constraints

Profit floor: $1,750

ROI floor: $1,073

Binding: ROI floor

Snapshot blind/as-is max hammer
$1,000
Modeled result at recorded bid
No bid recorded

This is a reseller ceiling, not a prediction of the auction's closing price. Any buyer premium or other unquoted acquisition charge must be deducted before bidding.

Special-protein workflow completeness and blind economics

The direct Type 864/2019 identity and visible internal hardware support a specific analyzer model, but neither a display-like peripheral nor a seller operational statement proves a usable control environment, complete peripherals, or application-level performance.

Exact-model asking evidence supports a specialist market but not a realized resale value. The blind model reserves separately for controller/software recovery, diagnostics, freight, repair, and liquidity risk.

  • Inventory the controller, software environment, licenses, interfaces, tote contents, reservoirs, cables, and peripherals before increasing the resale basis.
  • Demonstrate initialization, robotics, fluidics, temperature control, optics, barcode handling, and a meaningful QC/application sequence before representing the analyzer as working.

Recommendation and rationale

Snapshot decision

REVIEW

REVIEW: the exact 2019 Type 864 identity and seller-reported operation are useful, but independent functional proof, controller/software completeness, QC/service history, and realized-market evidence remain unresolved; use the $1,000 ceiling only as a blind/as-is threshold.

Confidence

Identity
High
Configuration
Medium
Condition
Low
Market Support
Medium
Costs
Medium Low
Why these confidence levels were assigned
Identity rationale
The manufacturer label directly establishes Optilite Type 864 and 2019 manufacture.
Configuration rationale
Visible robotics, fluid containers, cabling, and tote contents support partial completeness only; the controller/software environment remains unresolved.
Material unknowns: Unbounded
Condition rationale
Seller-reported operation is retained as seller evidence only; no independent functional, QC, calibration, or assay result is preserved.
Material unknowns: Unbounded
Market Support rationale
Exact-model asking evidence supports market existence but not realized resale.
Material unknowns: Bounded
Costs rationale
Palletized removal is bounded while controller/software, diagnostics, repair, and buyer-specific freight remain unquoted.
Material unknowns: Unbounded

Market support assesses how well the available evidence supports the resale assumption. Exact third-party marketplace records may be omitted when source-use rights are unclear. Each dimension uses the published methodology rubric; a confidence label measures the strength and boundedness of its evidence, not the likelihood of a profitable outcome.

Risks and unknowns

High · Controller Software Unresolved

A display-like peripheral does not establish a compatible controller, usable Optilite software, entitlement, or complete workflow.

Those dependencies control resale as a functioning analyzer.

High · Special Protein Function Unverified

Seller-reported operation does not establish robotics, fluidics, optics, QC, calibration, or assay performance.

Clinical-laboratory function cannot be inferred from exterior or internal assembly.

Sources

View cited public sources