Analysis snapshotSnapshot decision:REVIEW

Beckman Coulter Navios 10-Color/3-Laser Flow Cytometer — GSA Auction Analysis

A seller-confirmed Beckman Coulter Navios 10-color/3-laser system has palletized accessories and visible support hardware, but controller/software, complete inventory, startup, fluidics, laser, detector, acquisition, and QC evidence are unresolved. The $7,000 conservative resale model supports REVIEW and an $800 blind ceiling.

Analysis snapshot · Historical — observed September 16, 2026. The scheduled close date of Sep 17, 2026, 11:59 PM UTC has passed. The price, evidence, economics, and recommendation remain exactly as recorded and are not current.

Decision at a glance

RecommendationREVIEW
Maximum hammer bid
$800

The auction bid ceiling supported by this analysis, if one is defensible; see costs below for buyer-premium treatment.

Why this decision

REVIEW: the exact 10-color/3-laser identity and visible system hardware are promising, but repair-risk wording, controller/software uncertainty, unproven core function, and ask-heavy market evidence limit the blind/as-is maximum hammer to $800.

What to verify before bidding

  • A possible monitor or boxed accessories do not establish a compatible controller, acquisition software, license, or interfaces.
  • No core laser, fluidics, detector, sample-acquisition, or QC evidence is preserved.
Review all risks and unknowns

See costs and maximum bid for the displayed assumptions and buyer-premium treatment.

Snapshot auction result
Not captured
Bidders
Not captured
Blind-model resale
$7,000
Snapshot provenance and timing · observed Sep 16, 2026, 7:56 AM UTC · scheduled close Sep 17, 2026, 11:59 PM UTC

GSA Auctions sale 1-1-QSC-I-26-326, lot 022 · Louis Stokes VA Medical Center, Cleveland, Ohio
Snapshot observed Sep 16, 2026, 7:56 AM UTC
Scheduled close Sep 17, 2026, 11:59 PM UTC
Analysis prepared Sep 16, 2026, 7:56 AM UTC
Published September 16, 2026 · page updated September 16, 2026

Prepared and published by Oleksii Udod for LotCaliper.

View official GSA lot

Continue after this auction

This auction snapshot is historical. Continue with related research, or review a relevant published analysis whose recorded scheduled close has not passed.

These are individually prepared, time-stamped analyses with recorded scheduled closes. Verify the official auction before acting.

Auction snapshot

This page preserves the September 16, 2026 GSA open-auction observation for sale 1-1-QSC-I-26-326, lot 022, item 377480. No numeric high bid or bidder count was displayed; blank fields are not a $0 result.

A desirable flow-cytometry configuration is not a complete working system until its controller/software, fluidics, lasers, detectors, acquisition path, and QC evidence are established.

Review and correction history

Independent review completed. No corrections recorded.

View review and correction details

Independent review completed

Status: Approved · Reviewed September 16, 2026

A source-first second-pass review accepted the conservative valuation and ceiling without treating seller or photo evidence as functional proof.

How independent review works
How evidence is classified

Seller facts, photo-derived observations, manufacturer context, unknowns, and LotCaliper assumptions remain explicitly separated.

Source Confirmed

Seller-confirmed

  • Seller-confirmed Navios 10-color/3-laser identity, used/boxed condition, accessory packaging, and repair-risk wording.

Photo Derived

Photo-derived

  • The main Navios instrument, support/fluidics hardware, waste container, carousel-like components, and palletized Beckman Coulter cartons are visible.

Not Established

Not established

  • Controller PC, software/license, complete accessory inventory, startup, fluidics, laser/detector function, sample acquisition, QC, service history, and freight/loading cost.

Manufacturer Context

Independent context

  • Beckman Coulter continues to publish Navios platform and security/support context, which does not establish service eligibility for this unit.

Lotcaliper Assumption

LotCaliper assumptions

  • $7,000 conservative resale after inventory and basic diagnostic work, without validated clinical performance or transferable software.
  • $800 blind maximum hammer after selling friction, logistics, diagnostics, configuration, uncertainty, and required-profit reserves.
Structured photo observations (3)

These are bounded observations from the preserved photographs. They do not establish functional operation or seller inclusion beyond the corresponding source evidence.

Model · Confirmed

Navios badge on the main instrument

Photo confidence: High

Module Accessory · Confirmed

Beckman Coulter-branded cartons strapped to pallets

Photo confidence: High

Module Accessory · Probable

Circular carousel-like sample-handling components

Photo confidence: Medium High

Equipment identity

Laboratory · quantity 1

Beckman Coulter Navios 10-color/3-laser flow cytometer

Identity confidence: High

Identity evidence
  • The seller identifies one Navios 10 Color/3 Laser system; supplied photographs show the Navios instrument and Beckman Coulter-branded packaging.

    GSA Auctions · Text · observed Sep 16, 2026, 7:56 AM UTC

  • Supplied photographs show the main Navios instrument badge.

    GSA Auctions · Image · observed Sep 16, 2026, 7:56 AM UTC

Configuration

Support/fluidics-cart assembly with visible waste hardware

Possible · Medium High confidence

Configuration evidence
  • A separate wheeled support/fluidics assembly and labeled waste container are visible, but no readable model plate establishes exact configuration.

    GSA Auctions · Image · observed Sep 16, 2026, 7:56 AM UTC

Complete boxed accessory, carousel, cable, and peripheral inventory

Not established · no source evidence establishes presence or absence

Configuration evidence
  • Boxed Beckman Coulter accessories and carousel-like components are visible but not inventoried sufficiently to prove a complete system.

    GSA Auctions · Image · observed Sep 16, 2026, 7:56 AM UTC

Compatible controller PC, Navios acquisition software, transferable license, storage media, and interface cables

Not established · no source evidence establishes presence or absence

Configuration evidence
  • A possible display peripheral does not establish a compatible controller, usable software environment, license, or required interfaces.

    GSA Auctions · Image · observed Sep 16, 2026, 7:56 AM UTC

Condition

As-is / repair risk

The seller states the used, boxed system may have missing parts or require repairs, including unknown repairs.

Evidence confidence: High

Condition evidence
  • Seller condition text preserves missing-parts and repair-risk wording.

    GSA Auctions · Text · observed Sep 16, 2026, 7:56 AM UTC

Untested

No startup, self-test, laser, fluidics, detector, carousel-loader, sample-acquisition, or valid QC result is demonstrated.

Evidence confidence: High

Condition evidence
  • Supplied photographs establish physical presence and palletization, not functional or QC performance.

    GSA Auctions · Image · observed Sep 16, 2026, 7:56 AM UTC

Market evidence

Market evidence strength

Medium Low

Public auditability

Full

Evidence recency

Mixed

Current exact-model dealer asking and wanted evidence, plus a non-realized exact-model priced reference, support specialist liquidity but not a qualified realized resale basis.

Model matchEvidence typeTreatmentCount
Exact ModelAskingContext Only1
Exact ModelEnded Non RealizedUsed In Valuation1
Exact ModelInsufficient EvidenceContext Only1
SourceEvidenceObservedPriceRelevance and treatment
LabMakelaar / Machinio Navios

Dealer

Asking

Exact Model

September 16, 2026$17,347

Used dealer system with PC/software and refurbishing service; stronger configuration than the GSA lot.

Context Only

Industrial Lynx Navios

Dealer

Ended Non Realized

Exact Model

September 16, 2026$8,100

Out-of-stock inspected/fully functional system with compressor, cables, and six-month warranty.

Used In Valuation

DOTmed Navios wanted listing

Marketplace

Insufficient Evidence

Exact Model

September 16, 2026Not published

Wanted listing seeks complete good-working 3-laser/10-color systems; liquidity evidence without a public price.

Context Only

Sensitivity and robustness

Requires Verification

The scenarios show potentially workable economics, but the stated verification gate must be resolved before bidding responsibly.

Verified complete system

Max hammer
$4,500
Gross resale model
$10,000–$15,000

A conditional ceiling near $4,500 requires included controller/software, normal startup/QC, lower configuration and repair reserves, and a materially stronger resale basis.

Costs and maximum bid

Transaction / operating costAmountBasis
Selling Fees$1,050LotCaliper assumption: 15% of modeled gross resale.
Shipping$1,000LotCaliper pickup, packing, rigging, and inbound-logistics reserve; not a carrier or GSA quote.
Testing Repair Reserve$1,000LotCaliper functional-test and diagnostic reserve.
Risk Buffer$750Unresolved controller/software, functional/QC, repair, removal, and specialist-liquidity risk reserve.
Other$1,000LotCaliper missing-component, configuration, or recovery reserve; not a seller or service quote.
Buyer premium
Not modeled
Return requirements
$500 minimum profit · 30% minimum ROI
Computed bid constraints

Profit floor: $1,700

ROI floor: $827

Binding: ROI floor

Snapshot blind/as-is max hammer
$800
Modeled result at recorded bid
No bid recorded

This is a reseller ceiling, not a prediction of the auction's closing price. Any buyer premium or other unquoted acquisition charge must be deducted before bidding.

Flow-cytometry workflow completeness and blind economics

The 10-color/3-laser identity and visible support hardware establish more than a bare chassis, but a possible display and boxed accessories do not prove a transferable acquisition environment or complete system inventory.

Seller repair-risk wording and no demonstrated startup, fluidics, laser, detector, acquisition, or QC result require a blind/as-is model. Current dealer asks indicate specialist interest but are not realized resale evidence.

  • Inventory controller, software/license, interfaces, fluidics, sample-handling components, and boxed accessories before increasing the resale basis.
  • Demonstrate startup, normal fluidics and laser/detector behavior, sample acquisition, and current QC before representing the system as working.

Recommendation and rationale

Snapshot decision

REVIEW

REVIEW: the exact 10-color/3-laser identity and visible system hardware are promising, but repair-risk wording, controller/software uncertainty, unproven core function, and ask-heavy market evidence limit the blind/as-is maximum hammer to $800.

Confidence

Identity
High
Configuration
Medium
Condition
Low
Market Support
Medium Low
Costs
Medium Low
Why these confidence levels were assigned
Identity rationale
Seller-confirmed 10-color/3-laser identity and visible Navios equipment support the model identity.
Configuration rationale
Main system, support hardware, cartons, and carousel-like components are visible, while controller/software and complete inventory remain unresolved.
Material unknowns: Unbounded
Condition rationale
Seller repair-risk wording and no demonstrated startup, acquisition, or QC leave the functional state unresolved.
Material unknowns: Unbounded
Market Support rationale
Exact-model asking and wanted evidence supports a specialist market without establishing realized value.
Material unknowns: Bounded
Costs rationale
Palletization helps, but buyer-managed removal, diagnostics, controller/software recovery, and repair remain unquoted.
Material unknowns: Unbounded

Market support assesses how well the available evidence supports the resale assumption. Exact third-party marketplace records may be omitted when source-use rights are unclear. Each dimension uses the published methodology rubric; a confidence label measures the strength and boundedness of its evidence, not the likelihood of a profitable outcome.

Risks and unknowns

High · Controller Software Unresolved

A possible monitor or boxed accessories do not establish a compatible controller, acquisition software, license, or interfaces.

Those dependencies control whether this is a usable flow-cytometry system.

High · Flow Cytometry Function Unverified

No core laser, fluidics, detector, sample-acquisition, or QC evidence is preserved.

Exterior condition cannot establish analytical performance.