The matched microscope/controller relationship is positive evidence because an unrelated controller would create a serious compatibility risk. It is not evidence of working laser capture, vacuum, image acquisition, or a usable legacy computer path.
The original architecture depended on a dedicated computer, frame-grabber/video hardware, serial control, and Arc2.0.0 software. None is established in the lot, and optional fluorescence should not be inferred from platform capability.
- Demonstrate vacuum, stage/cap movement, laser aiming/pulse, camera image, and actual transfer to a CapSure film before assigning working-system value.
- Confirm whether any original workstation, monitor, frame-grabber, Arc2.0.0 installation, backups, or fluorescence hardware is stored separately and included.