Analysis snapshotSnapshot decision:REVIEW

Bruker Ultraflex-Family MALDI-TOF/TOF Mass Spectrometer — Research Lab Equipment Lot

A Bruker Ultraflex-family MALDI-TOF/TOF system is photo-confirmed in this mixed research-lab lot, but exact generation, workstation/software, vacuum, laser, high-voltage, detector, and removal state are unresolved. The $3,500 conservative system-only model supports REVIEW and a ~$650 blind ceiling.

Analysis snapshot · Historical — observed September 15, 2026. The scheduled close date of September 17, 2026 has passed. The price, evidence, economics, and recommendation remain exactly as recorded and are not current.

Decision at a glance

RecommendationREVIEW
Maximum hammer bid
$650

The auction bid ceiling supported by this analysis, if one is defensible; see costs below for buyer-premium treatment.

Why this decision

REVIEW: the mixed lot contains a defensible Ultraflex-family system identity, but exact configuration, function, and heavy removal remain too unresolved for a BUY; use the ~$650 ceiling only as a blind/as-is threshold.

What to verify before bidding

  • Ultraflex II-generation appearance does not establish an exact model, suffix, compatibility, or value-bearing options.
  • Workstation/software, vacuum hardware, targets, pumps, and accessories are not established.
  • No vacuum, laser, high-voltage, detector, or acquired-spectrum performance is demonstrated.
Review all risks and unknowns

See costs and maximum bid for the displayed assumptions and buyer-premium treatment.

Snapshot auction result
Not captured
Bidders
Not captured
Blind-model resale
$3,500
Snapshot provenance and timing · observed September 15, 2026 · scheduled close September 17, 2026

GSA Auctions sale Not captured, lot Not captured · Location not captured in the supplied evidence
Snapshot observed September 15, 2026
Scheduled close September 17, 2026
Analysis prepared September 15, 2026
Published September 15, 2026 · page updated September 15, 2026

Prepared and published by Oleksii Udod for LotCaliper.

View official GSA lot

Continue after this auction

This auction snapshot is historical. Continue with related research, or review a relevant published analysis whose recorded scheduled close has not passed.

These are individually prepared, time-stamped analyses with recorded scheduled closes. Verify the official auction before acting.

Auction snapshot

This page preserves the September 15, 2026 GSA open-auction observation for item 377412. No observed bid or bidder count is preserved; blank fields are not a $0 result or later auction outcome.

Family-level identity is only the first step for a legacy MALDI system: workstation/software, vacuum, laser, detector, serviceability, and removal can control the bid ceiling.

Review and correction history

Independent review not required. No corrections recorded.

View review and correction details

Independent review not required

No mandatory independent review was required under the current gate. This is not an approval.

How independent review works
How evidence is classified

Seller facts, photo-derived observations, manufacturer context, unknowns, and LotCaliper assumptions remain explicitly separated.

Source Confirmed

Seller-confirmed

  • A Bruker Ultraflex-family MALDI-TOF/TOF system, laser warning, sample-loading area, and substantial exterior/electronics sections are photo-confirmed.

Photo Derived

Photo-derived

  • The chassis appears substantially assembled without obvious severe exterior deformation; that is not operating evidence.

Not Established

Not established

  • Exact generation/suffix, workstation/software/licenses, vacuum and pump configuration, targets, laser, high-voltage, detector, acquired spectrum, secondary equipment identity, and removal cost.

Manufacturer Context

Independent context

  • Bruker lists ultraflex II and III as end-of-life platforms; specialist support remains visible but does not establish serviceability for this unit.

Lotcaliper Assumption

LotCaliper assumptions

  • $3,500 conservative whole-lot resale dominated by the untested Bruker system, with no base value for secondary equipment.
  • $650 blind maximum hammer after selling, logistics, diagnostics, uncertainty, and required-profit reserves.

Equipment identity

Mass Spectrometry · quantity 1

Bruker Ultraflex-family MALDI-TOF/TOF mass spectrometer

Identity confidence: Medium High

Identity evidence
  • Supplied photographs show Bruker branding, a laser warning, sample-loading geometry, and system architecture consistent with the Ultraflex MALDI-TOF/TOF family.

    GSA Auctions · Image · observed September 15, 2026

Configuration

Exact Ultraflex generation, suffix, factory options, laser generation, and detector configuration

Possible · Medium High confidence

Configuration evidence
  • The enclosure is consistent with an Ultraflex II-generation system, but no readable model or serial plate establishes the exact generation or options.

    GSA Auctions · Image · observed September 15, 2026

Compatible workstation, flexControl/flexAnalysis environment, software media, licenses, and security devices

Not established · no source evidence establishes presence or absence

Configuration evidence
  • The supplied photographs do not establish a transferable computer, software, license, or control environment.

    GSA Auctions · Image · observed September 15, 2026

Vacuum pumps, target plates, calibration standards, ancillary controllers, cables, and complete MALDI workflow

Not established · no source evidence establishes presence or absence

Configuration evidence
  • Exterior views do not establish the value-bearing vacuum, target, pump, controller, or accessory configuration.

    GSA Auctions · Image · observed September 15, 2026

Additional secondary laboratory equipment in the mixed lot

Not established · no source evidence establishes presence or absence

Configuration evidence
  • Additional laboratory hardware is visible but not identified, configured, or condition-qualified sufficiently for value credit.

    GSA Auctions · Image · observed September 15, 2026

Condition

Untested

No power-on, self-test, achieved vacuum, laser/interlock, high-voltage, detector, target movement, workstation communication, or acquired-spectrum evidence is supplied.

Evidence confidence: High

Condition evidence
  • Substantial exterior assembly is visible, but no operating screen or functional result is preserved.

    GSA Auctions · Image · observed September 15, 2026

Market evidence

Market evidence strength

Medium

Public auditability

Full

Evidence recency

Mixed

Public exact and related Ultraflex-family asks and parts evidence support residual value, but realized-sale evidence is thin and the GSA submodel and function remain unresolved.

Model matchEvidence typeTreatmentCount
Exact ModelAskingUsed In Valuation1
Exact ModelAskingContext Only1
Related ModelAskingExcluded2
Related ModelInsufficient EvidenceContext Only1
SourceEvidenceObservedPriceRelevance and treatment
Machinio Ultraflex II TOF/TOF 200

Dealer

Asking

Exact Model

September 15, 2026$4,500

Removed from an operating environment but not fully tested; offered parts/as-is.

Used In Valuation

Lab Assets Ultraflex 2

Dealer

Asking

Exact Model

September 15, 2026£5,999

Used/fully operational specialist listing; stronger function than the GSA evidence.

Context Only

Machineseeker Ultraflex TOF/TOF UT0100

Dealer

Asking

Related Model

September 15, 2026€25,000

Excellent-condition related-family system; materially stronger than the GSA lot.

Excluded — Condition and configuration are too strong for the conservative base valuation.

Lab Liquidators Ultraflex II vacuum chamber

Dealer

Asking

Related Model

September 15, 2026$699

Internal vacuum component only; evidence of a parts ecosystem, not a whole system.

Excluded — Component-only pricing cannot value the complete instrument.

UniGreen Scheme Ultraflex TOF/TOF

Other

Insufficient Evidence

Related Model

September 15, 2026Not published

Quote-only spares/repair system with intermittent 30 kV and laser faults; 500 kg handling context.

Context Only

Sensitivity and robustness

Requires Verification

The scenarios show potentially workable economics, but the stated verification gate must be resolved before bidding responsibly.

Untested family-level system

Max hammer
$650
Gross resale model
$2,500–$5,000

At $3,000 conservative resale, the rounded ROI ceiling is about $350; with $1,500 logistics, it is about $150. No value is assigned to unidentified secondary equipment.

Costs and maximum bid

Transaction / operating costAmountBasis
Selling Fees$525LotCaliper assumption: 15% of modeled gross resale.
Shipping$1,000LotCaliper pickup, packing, rigging, and inbound-logistics reserve; not a carrier or GSA quote.
Testing Repair Reserve$250LotCaliper functional-test and diagnostic reserve.
Risk Buffer$350Unresolved exact generation, workstation/software, vacuum, laser, high-voltage, detector, heavy removal, and specialist-liquidity risk.
Buyer premium
Not modeled
Return requirements
$500 minimum profit · 30% minimum ROI
Computed bid constraints

Profit floor: $875

ROI floor: $688

Binding: ROI floor

Snapshot blind/as-is max hammer
$650
Modeled result at recorded bid
No bid recorded

This is a reseller ceiling, not a prediction of the auction's closing price. Any buyer premium or other unquoted acquisition charge must be deducted before bidding.

Family identity, workflow completeness, and heavy-logistics economics

The public identity stays at Bruker Ultraflex-family MALDI-TOF/TOF. An Ultraflex II-generation interpretation is possible, not confirmed, so no exact-generation premium is modeled.

Public market asks show residual specialist and parts value, but asking prices are not realized sales and do not resolve the GSA system's untested vacuum, laser, high-voltage, detector, or workstation state.

  • Obtain a readable model plate and inventory workstation, software/licenses, pumps, targets, controllers, cables, and accessories before raising the value basis.
  • Demonstrate achieved vacuum, normal laser/interlocks and high voltage, detector response, workstation communication, and a plausible calibration spectrum before treating the system as working.

Recommendation and rationale

Snapshot decision

REVIEW

REVIEW: the mixed lot contains a defensible Ultraflex-family system identity, but exact configuration, function, and heavy removal remain too unresolved for a BUY; use the ~$650 ceiling only as a blind/as-is threshold.

Confidence

Identity
Medium High
Configuration
Low Medium
Condition
Low
Market Support
Medium
Costs
Medium Low
Why these confidence levels were assigned
Identity rationale
Bruker branding and system architecture support the Ultraflex family, while exact generation remains unresolved.
Configuration rationale
The workstation, software, vacuum, targets, pumps, controllers, and accessory configuration are not established.
Material unknowns: Unbounded
Condition rationale
No function is demonstrated across the system's decision-critical vacuum, laser, high-voltage, detector, and acquisition path.
Material unknowns: Unbounded
Market Support rationale
Current exact/related asking evidence and a parts market support residual value but not an exact realized value.
Costs rationale
400–500 kg-class handling is documented in market context, while actual removal and freight remain unquoted.
Material unknowns: Unbounded

Market support assesses how well the available evidence supports the resale assumption. Exact third-party marketplace records may be omitted when source-use rights are unclear. Each dimension uses the published methodology rubric; a confidence label measures the strength and boundedness of its evidence, not the likelihood of a profitable outcome.

Risks and unknowns

High · Exact Generation Unresolved

Ultraflex II-generation appearance does not establish an exact model, suffix, compatibility, or value-bearing options.

No readable model or serial plate is preserved.

High · Software And Workflow Unresolved

Workstation/software, vacuum hardware, targets, pumps, and accessories are not established.

The analyzer chassis alone does not prove a usable MALDI workflow.

High · Functionality Unverified

No vacuum, laser, high-voltage, detector, or acquired-spectrum performance is demonstrated.

These are the decision-critical failure modes for an older MALDI system.